HFCN Test Fixture,High frequency coupling network test fixture. Compliant with IEC 60601-2-25, IEC 60601-2-27, China YY1079, and ANSI / AAMI EC13.

  • 产品型号:
  • 厂商性质:代理商
  • 更新时间:2023-12-04
  • 访  问  量:196
立即咨询

联系电话:

详细介绍

HFCN Test Fixture,High frequency coupling network test fixture. Compliant with IEC 60601-2-25, IEC 60601-2-27, China YY1079, and ANSI / AAMI EC13.

OVERVIEW

Elaborated design offers convenience and safety

Dual module design supports different standard requirements

Built-in impedance 51k/47nF for skin impedance simulation

Simulate 1mV ECG signal for electrosurgical interference suppression (ESIS)

Power rating >300W and resistor inductance <5μH

Able to use BB-C1 breakout box bundled with SECG 4.0 and MECG 2.0

SPECIFICATION

ItemsSpecifications

Standard CoverageBuilt-in test circuits as defined in IEC 60601-2-25/-2-27/-2-49:2011 Clause 202.6.2.101 (Electrosurgeryinterference); China YY1079:2008 Clause 5.2.8.13; AAMI EC13:2002 Clause 5.2.9.14 ESIO/ESIS tests

Test Fixture Power Rating>300W

Resistor Inductance<5μH

Frequency RangeDC to 100 MHz

Use Non-Inductive Power ResistorsResistance Tolerances to ±5%, TCR of ±100 ppm/C, Power Rating 200 W/pcs

Skin Impedance Simulated CircuitResister: 51 kΩ <±5%; Capacitor: 47 nF <±5% C0G

ECG Simulator in Module B(defined in EC13, acting on LL, V1~V6)

- Amplitude: 1 mV ±10%

- Pulse width: 75 ms ±10%

- Rate: 73 bpm ±5%

HFCN Test Fixture

638354679245575388228.jpg



留言询价

留言框

  • 产品:

  • 您的单位:

  • 您的姓名:

  • 联系电话:

  • 常用邮箱:

  • 省份:

  • 详细地址:

  • 补充说明:

  • 验证码:

    请输入计算结果(填写阿拉伯数字),如:三加四=7

C

CODE
关注我们
  • 联系电话:13671818969

  • 联系邮箱:lisa815@yeah.net

  • 公司地址:中国(上海)自由贸易试验区临港新片区老芦公路536号

Copyright © 2024 上海力孚同科技有限公司版权所有   备案号:   技术支持:化工仪器网

sitmap.xml   管理登陆